A semiautomated procedure for urinary D-glucaric acid using a centrifugal analyzer
Journal of Analytical Toxicology
|January 1, 1986
Summary
Quantifying D-glucaric acid using an enzyme inhibition method was improved. A semiautomated approach on a centrifugal analyzer increased throughput and showed high correlation with manual methods for measuring this important biomarker.
Area of Science:
- Biochemistry
- Clinical Chemistry
- Analytical Chemistry
Background:
- D-glucaric acid is a biomarker for metabolic processes.
- The enzyme inhibition method is standard for its quantification.
- This method is a two-step procedure involving beta-glucuronidase inhibition.
Purpose of the Study:
- To adapt the enzyme inhibition method for D-glucaric acid quantification to a Cobas-Bio centrifugal analyzer.
- To evaluate the performance of the semiautomated method compared to the manual method.
Main Methods:
- Urine was boiled at acidic pH to form 1,4-glucarolactone, inhibiting beta-glucuronidase.
- Beta-glucuronidase inhibition was measured as an indirect indicator of D-glucaric acid.
- The second step was adapted for a Cobas-Bio centrifugal analyzer.
Main Results:
- The semiautomated method showed high correlation with the manual method (r = 0.976).
- Precision was comparable between the manual and semiautomated methods.
- Sample throughput increased from 8 to 20 samples per day.
Conclusions:
- The semiautomated enzyme inhibition method is a reliable and more efficient alternative for quantifying D-glucaric acid.
- This adaptation enhances the practicality of measuring D-glucaric acid in clinical settings.