Behavior of leadless atrioventricular synchronous pacing during exercise
Christophe Garweg1, Thomas Van Weyenbergh1, Todd Sheldon2
1Department of Cardiovascular Sciences, University of Leuven, Leuven, Belgium; Cardiology, University Hospitals Leuven, Leuven, Belgium.
Background:
The leadless Micra AV pacemaker is designed to provide atrioventricular (AV) synchronous tracking by detecting atrial contraction. Detection of the mechanical atrial signals can become challenging at a fast sinus rate.
Objective:
The purpose of the study was to evaluate the AV synchronous performance during exercise in outpatients implanted with a Micra AV pacemaker.
Methods:
Patients were enrolled at least 1 month after Micra AV implantation and underwent a cycle test protocol. Serial device interrogations (each minute) and continuous electrocardiograms were collected to measure AV synchrony and determine maximum achieved sinus and ventricular rates for each patient. In addition, the A1, A2, A3, and A4 accelerometer signal amplitudes were measured at the start and peak of exercise.
Results:
Thirty-five patients (mean age 75.6 ± 13.4 years; 28 (80%) male) were enrolled in the study; 22 (64%) were predominantly ventricular paced (>90%) during exercise. Mean AV synchrony was 90.4% in the entire cohort and 84.7% in patients with high-degree AV block. The mean amplitude of the accelerometer signals increased significantly from the start to the peak of exercise: A1, 4.1-6.3 m/s2; A2, 2.4-3.8 m/s2; and A4, 4.5-7.6 m/s2 (P < .01 for all). The time from the ventricular ecvent (VS/VP) to A2 showed a decrease of 25 ms for each 100 ms decrease of the R-R interval.
Conclusion:
Maintaining AV synchrony during maximal exercise in elderly patients is achievable by adequate detection of atrial contraction at fast sinus rates by the leadless Micra AV pacemaker. All components of the accelerometer signal increased, likely because of increased contractility related to exercise.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
08:05Optimization of Transesophageal Atrial Pacing to Assess Atrial Fibrillation Susceptibility in Mice
Published on: June 29, 2022
Related Concept Videos
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Electrophysiology of Normal Cardiac Rhythm
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow...
The Cardiac Cycle
The Process
Electrical signals—sent from the sinoatrial (SA) node in the right atrial wall to the atrioventricular (AV) node between the right atrium and right ventricle—cause both atria to simultaneously contract. When the signal reaches the AV node, it pauses for approximately a tenth of a second, allowing the atria to contract and...
Heart Failure Drugs: Inotropic Agents
Exercise and Cardiac Output
Sustained exercise increases the muscles' oxygen demand, which can be...