Joint suppression of cardiac bSSFP cine banding and flow artifacts using twofold phase-cycling and a dual-encoder
Zhuo Chen1, Yiwen Gong2, Haiyang Chen1
1National Engineering Research Center of Advanced Magnetic Resonance Technologies for Diagnosis and Therapy, School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, China.
Background:
Cardiac balanced steady state free precession (bSSFP) cine imaging suffers from banding and flow artifacts induced by off-resonance. The work aimed to develop a twofold phase cycling sequence with a neural network-based reconstruction (2P-SSFP+Network) for a joint suppression of banding and flow artifacts in cardiac cine imaging.
Methods:
A dual-encoder neural network was trained on 1620 pairs of phase-cycled left ventricular (LV) cine images collected from 18 healthy subjects. Twenty healthy subjects and 25 patients were prospectively scanned using the proposed 2P-SSFP sequence. bSSFP cine of a single RF phase increment (1P-SSFP), bSSFP cine of a single radiofrequency (RF) phase increment with a network-based artifact reduction (1P-SSFP+Network), the averaging of the two phase-cycled images (2P-SSFP+Average), and the proposed method were mutually compared, in terms of artifact suppression performance in the LV, generalizability over altered scan parameters and scanners, suppression of large-area banding artifacts in the left atrium (LA), and accuracy of downstream segmentation tasks.
Results:
In the healthy subjects, 2P-SSFP+Network showed robust suppressions of artifacts across a range of phase combinations. Compared with 1P-SSFP and 2P-SSFP+Average, 2P-SSFP+Network improved banding artifacts (3.85 ± 0.67 and 4.50 ± 0.45 vs 5.00 ± 0.00, P < 0.01 and P = 0.02, respectively), flow artifacts (3.35 ± 0.78 and 2.10 ± 0.77 vs 4.90 ± 0.20, both P < 0.01), and overall image quality (3.25 ± 0.51 and 2.30 ± 0.60 vs 4.75 ± 0.25, both P < 0.01). 1P-SSFP+Network and 2P-SSFP+Network achieved a similar artifact suppression performance, yet the latter had fewer hallucinations (two-chamber, 4.25 ± 0.51 vs 4.85 ± 0.45, P = 0.04; four-chamber, 3.45 ± 1.21 vs 4.65 ± 0.50, P = 0.03; and left atrium (LA), 3.35 ± 1.00 vs 4.65 ± 0.45, P < 0.01). Furthermore, in the pulmonary veins and LA, 1P-SSFP+Network could not eliminate banding artifacts since they occupied a large area, whereas 2P-SSFP+Network reliably suppressed the artifacts. In the downstream automated myocardial segmentation task, 2P-SSFP+Network achieved more accurate segmentations than 1P-SSFP with different phase increments.
Conclusions:
2P-SSFP+Network jointly suppresses banding and flow artifacts while manifesting a good generalizability against variations of anatomy and scan parameters. It provides a feasible solution for robust suppression of the two types of artifacts in bSSFP cine imaging.