Photon-counting X-ray Computed Tomography Using a Cadmium Telluride Flat Panel Detector with High Spatial Resolutions
Jiro Sato1, Eiichi Sato2, Kazuki Ito1
1Department of Surgery, Toho University Ohashi Medical Center, Meguro, Tokyo, Japan.
Introduction:
To perform energy-dispersive X-ray imaging, we constructed a photon-counting X-ray computed tomography (CT) scanner to perform enhanced K-edge CT.
Methods:
X-ray photons penetrating through an object were detected using a cadmium telluride flat panel detector (FPD) with pixel dimensions of 100 × 100 mm2, and 720 radiograms from the FPD were sent to the personal computer to reconstruct tomograms. Gadolinium (Gd) K-edge energy is 50.2 keV, and Gd-Kedge CT was carried out using photons with an energy range of 50-100 keV.
Results:
Compared with low-energy CT of 15-50 keV, the gray density of muscle and bone substantially decreased, and the image contrast of Gd media was improved utilizing Gd-K-edge CT.
Conclusion:
Using the cone beam, the effective pixel dimensions were 80 × 80 μm2, and blood vessels were observed at a high contrast using Gd-Kedge CT.
More Related Videos
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
07:013D Imaging of Soft-Tissue Samples using an X-ray Specific Staining Method and Nanoscopic Computed Tomography
Published on: October 24, 2019
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
X-ray Imaging