Comparing the Performance of Scatter Correction Methods in Cardiac SPECT Imaging with Technetium-99m and Thallium-201
Mahsa Noori-Asl1, Maryam Eghbal1
1Department of Physics, Faculty of Science, University of Mohaghegh Ardabili, Ardabil, Iran.
Purpose:
This study aims to evaluate the performance of dual-energy window (DEW) and triple-energy window (TEW) scatter correction methods in cardiac SPECT imaging with technetium-99m (Tc-99m) and thallium-201 (Tl-201) radioisotopes.
Materials And Methods:
The SIMIND Monte Carlo program was used to simulate the imaging system and produce the required projections. Two phantoms, including the simple cardiac phantom and the NCAT phantom, were used to evaluate the scatter correction methods. The simulations were repeated 5 times for each phantom and finally, the mean values obtained from these 5 tests were used in the analysis of the results.
Results:
The obtained results from this study show that in the case of both investigated phantoms, the use of correction methods leads to improve the contrast of the images obtained from Tc-99m and Tl-201 radioisotopes. In the case of the simple cardiac phantom, the use of DEW and TEW correction methods leads to a relative increase in image contrast of about 23.88% and 12.23% for 99mTc radioisotope and about 29.19% and 20.98% for 201Tl radioisotope, respectively. This relative increase in the case of the NCAT phantom is about 22.48% and 19.43% for 99mTc radioisotope and about 27.74% and 24.74% for 201Tl radioisotope, respectively.
Conclusion:
According to the obtained results, despite the higher contrast of the noncorrected images of 99mTc radioisotope, the relative increase in contrast of the corrected images of 201Tl radioisotope is more than that of 99mTc radioisotope. Furthermore, for both radioisotopes, the relative increase related to the DEW method is higher than the TEW method.
More Related Videos
06:29Cardiac Magnetic Resonance for the Evaluation of Suspected Cardiac Thrombus: Conventional and Emerging Techniques
Published on: June 11, 2019
09:07Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System V: CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...