Purpose:
Blood is commonly treated as single-phase homogeneous fluid in numerical simulations of blood flow within fiber bundles of blood oxygenators. However, microfluidics tests revealed the presence of hematocrit heterogeneity in blood flowing across such geometries. Given the significant role of red blood cells (RBCs) in the oxygenation process, this study aims to propose a multiphase blood model able to correctly describe the experimental evidence and computationally investigate hematocrit heterogeneities inside fiber bundles.
Methods:
The experimental results of microfluidics tests performed in a previous study were processed and based on quantitative data of image intensity, a two-phase blood model following the Eulerian-Eulerian approach was calibrated and evaluated in its predictive ability against the experimental data. The two-phase model was then used to study the RBCs distribution inside different fiber bundles at average hematocrit values of 25% and 35%, representative of hemodilution in extracorporeal blood circulation.
Results:
The numerical model proved to be able to describe and predict the experimental phase separation between plasma and RBCs within the microchannel geometry at different test conditions. Moreover, blood flow simulation in commercial fiber bundles revealed the presence of specific patterns in hematocrit distribution and their dependence on variations in bundle microstructure.
Conclusion:
The two-phase blood model proposed in this study provides a tool for advanced evaluation of local fluid dynamics and identification of optimal bundle microstructure allowing further gas transfer simulations to account for a reliable heterogeneous distribution of RBCs around the oxygenating fibers.
More Related Videos
09:38A Microfluidic Flow Chamber Model for Platelet Transfusion and Hemostasis Measures Platelet Deposition and Fibrin Formation in Real-time
Published on: February 14, 2017
08:44An In Vitro Hemodynamic Loop Model to Investigate the Hemocytocompatibility and Host Cell Activation of Vascular Medical Devices
Published on: August 21, 2020
Related Concept Videos
Blood Flow
Compartment Models: Two-Compartment Model
Oxygen Transport in the Blood
Physiological Pharmacokinetic Models: Blood Flow-Limited Versus Diffusion-Limited Models