Continuous vital sign monitoring on surgical wards: The COSMOS pilot
Nikola Anusic1, Alper Gulluoglu1, Elyad Ekrami1
1Research Fellow, OutcomesResearch Consortium, Department of Anesthesiology, Cleveland Clinic, Cleveland, OH, USA.
Study Objectives:
Alerts for vital sign abnormalities seek to identify meaningful patient instability while limiting alarm fatigue. Optimal vital sign alarm settings for postoperative patients remain unknown, as is whether alerts lead to effective clinical responses reducing vital sign disturbances. We conducted a 2-phase pilot study to identify thresholds and delays and test the hypothesis that alerts from continuous monitoring reduce the duration of vital sign abnormalities.
Design:
Two-phase pilot.
Patients:
250 adults having major non-cardiac surgery.
Setting:
Surgical wards.
Intervention:
All patients had routine vital sign monitoring by nurses at 4-h intervals. We initially continuously recorded clinician-blinded saturation, heart rate, and respiratory rate in 100 patients. In the second phase, we randomized 150 patients to blinded versus unblinded continuous vital sign monitoring. In unblinded patients, nurses were verbally alerted to abnormal vital signs.
Measurements:
In the first phase, we modeled expected alarm counts using 6082 h of continuous oxygen saturation, heart rate, and respiratory rate data. Thresholds and delays targeting ∼3 alarms per patient per day were selected for phase two. Primary analysis assessed the effect of unblinded monitoring across a 5-component primary composite of cumulative durations of vital sign abnormalities. Secondary outcomes included fraction of alerts deemed meaningful by nurses and number of clinical interventions.
Results:
In phase one, we identified alarm settings that yielded an average of 2.3 alerts per patient per day. In phase two, the average relative effect ratio of geometric duration means for vital signs exceeding thresholds was 0.75 [95 % CI: 0.51, 1.1], P = 0.17. Sixty alarms (82 %) were deemed useful in unblinded patients, leading to 60 % more interventions in unblinded patients.
Conclusions:
We were able to select continuous saturation, heart rate, and respiratory rate thresholds that generated about 2 alerts per patient per day, nearly all of which were considered useful by nurses. Unblinded monitoring and nursing alerts led to interventions (mostly increasing oxygen delivery) that non-significantly reduced vital sign abnormalities by 25 %.
Clinicaltrials:
gov registration: NCT05280574.
More Related Videos
10:38Observational Study Protocol for Repeated Clinical Examination and Critical Care Ultrasonography Within the Simple Intensive Care Studies
Published on: January 16, 2019
05:01A Detailed Protocol for Physiological Parameters Acquisition and Analysis in Neurosurgical Critical Patients
Published on: October 17, 2017
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Introduction to Vital Signs
Vital signs help healthcare professionals assess an individual's well-being and detect any functional changes...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...