Open-chest cardiac ultrasound-mediated imaging with a vacuum coupler
Nilesh Mathuria1, Krithik Vishwanath2, Giorgio Brero3,4
1Houston Methodist Heart and Vascular Center, Houston Methodist Research Institute, Houston Methodist Hospital, Houston, Texas, USA.
Background:
A fundamental obstacle for the preclinical development of ultrasound-(US) mediated cardiac imaging remains cardiac motion, which limits interframe correlation during extended acquisition periods.
Purpose:
To address this need, we present the design and implementation of a 3D-printed vacuum coupler that stabilizes a US transducer on the epicardial surface of the heart for feasibility assessment and development of advanced, cardiac, US-mediated imaging approaches.
Methods:
The vacuum coupler was 3D printed with biocompatible resins and secured with a standard intraoperative suction aspirator. US-mediated imaging (i.e., B-mode and photoacoustic [PA] imaging) was performed in an open-chest porcine model with and without the vacuum coupler. Based on inter-frame displacement tracking and cross-correlation (CC) coefficients, changes in frame motion and stability were compared for each imaging mode/configuration through a prolonged (∼1 min) acquisition, while the impact on PA-based SO2 accuracy was assessed.
Results:
When compared to conventional handheld imaging, stand-off imaging, and coupler without suction, epicardial imaging with the vacuum coupler and suction applied led to a significantly reduced mean axial displacement of 0.15 mm versus 0.89, 0.49, & 0.49 mm, respectively (p-values ≤ 8.65e-7). Comparing the coupler without suction to that with suction applied, physiologically unrealistic SO2 estimates reduced from 1.72 to 0.81%, respectively, and lateral interframe displacement reduced from 4.58 to 2.01 mm, respectively (p-value = 5.07e-23). Overall, reduced cardiac tissue motion and increased interframe CC coefficient (baseline = 0.43 vs. coupler with suction = 0.80) allow for more accurate PA unmixing.
Conclusions:
Epicardial US-mediated imaging with a vacuum coupler reduces cardiac motion artifact, providing a consistent sampling of an intended region of interest (ROI) over multiple cardiac cycles. This could help facilitate the development of advanced US-mediated imaging, which is often hindered by cardiac motion. Stable implementation of these imaging techniques could allow for intra-operative assessments of local cardiac perfusion as well as tissue characterization.
More Related Videos
08:37Echocardiographic Assessment Using Subxiphoid-Only Examination for Hypotensive Patients
Published on: April 18, 2025
10:38Observational Study Protocol for Repeated Clinical Examination and Critical Care Ultrasonography Within the Simple Intensive Care Studies
Published on: January 16, 2019
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Imaging Studies for Cardiovascular System V: CT