Simultaneous multislice cardiac multimapping based on locally low-rank and sparsity constraints
Yixin Emu1, Yinyin Chen2, Zhuo Chen1
1National Engineering Research Center of Advanced Magnetic Resonance Technologies for Diagnosis and Therapy, School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, China.
Background:
Although quantitative myocardial T1 and T2 mappings are clinically used to evaluate myocardial diseases, their application needs a minimum of six breath-holds to cover three short-axis slices. The purpose of this work is to simultaneously quantify multislice myocardial T1 and T2 across three short-axis slices in one breath-hold by combining simultaneous multislice (SMS) with multimapping.
Methods:
An SMS-Multimapping sequence with multiband radiofrequency (RF) excitations and Cartesian fast low-angle shot readouts was developed for data acquisition. When 3 slices are simultaneously acquired, the acceleration rate is around 12-fold, causing a highly ill-conditioned reconstruction problem. To mitigate image artifacts and noise caused by the ill-conditioning, a reconstruction algorithm based on locally low-rank and sparsity (LLRS) constraints was developed. Validation was performed in phantoms and in vivo imaging, with 20 healthy subjects and 4 patients, regarding regional mean, precision, and scan-rescan reproducibility.
Results:
The phantom imaging shows that SMS-Multimapping with locally low-rank (LLRS) accurately reconstructed multislice T1 and T2 maps despite a six-fold acceleration of scan time. Healthy subject imaging shows that the proposed LLRS algorithm substantially improved image quality relative to split slice-generalized autocalibrating partially parallel acquisition. Compared with modified look-locker inversion recovery (MOLLI), SMS-Multimapping exhibited higher T1 mean (1118 ± 43 ms vs 1190 ± 49 ms, P < 0.01), lower precision (67 ± 17 ms vs 90 ± 17 ms, P < 0.01), and acceptable scan-rescan reproducibility measured by 2 scans 10-min apart (bias = 1.4 ms for MOLLI and 9.0 ms for SMS-Multimapping). Compared with balanced steady-state free precession (bSSFP) T2 mapping, SMS-Multimapping exhibited similar T2 mean (43.5 ± 3.3 ms vs 43.0 ± 3.5 ms, P = 0.64), similar precision (4.9 ± 2.1 ms vs 5.1 ± 1.0 ms, P = 0.93), and acceptable scan-rescan reproducibility (bias = 0.13 ms for bSSFP T2 mapping and 0.55 ms for SMS-Multimapping). In patients, SMS-Multimapping clearly showed the abnormality in a similar fashion as the reference methods despite using only one breath-hold.
Conclusion:
SMS-Multimapping with the proposed LLRS reconstruction can measure multislice T1 and T2 maps in one breath-hold with good accuracy, reasonable precision, and acceptable reproducibility, achieving a six-fold reduction of scan time and an improvement of patient comfort.