The impact of different prong-nares ratio on ventilation in COPD patients using nasal high-flow (NHF) - a
Jens Bräunlich1,2, Hubert Wirtz3
1Department of Respiratory Medicine, University of Leipzig, Liebigstrasse 20, 04103, Leipzig, Germany. highflow@web.de.
Introduction:
Nasal high flow (NHF) is a popular technique to provide support in respiratory failure in different conditions. Recently published bench studies have hypothesized that airway pressure can be increased by using different cannula sizes and corresponding prongs resulting in a range of prong-nare ratios. We conducted this study to verify these experimental findings in clinical practice.
Methods:
We characterized prong size and flow rate dependent changes in ventilation parameters and changes in hypercapnia in an interventional clinical setting. Outcome parameters included changes in mean airway pressure, tidal volume (TV), respiratory rate (RR), minute volume (MV) and decrease in pCO2. The ventilatory parameters were determined at 20, 30, 40 and 50 l/min with 3 different prong sizes. 20 and 40 l/min and the 3 different prong sizes were used to document the changes in pCO2.
Results:
In this study we demonstrate changes in ventilation with increasing flow rates of NHF. A significant increase in mean airway pressure was seen with every 10 l/min increase in flow rate. Respiratory rate and minute volume (using large prongs) changed significantly with larger increases in flow rate, while tidal volume was not significantly altered. When the flow rate was increased by 20 l/min (i.e. from 20 l/min to 40 l/min) capillary pCO2 decreased significantly. None of the measured values were significantly altered by the prong size used.
Conclusion:
In summary, we presented strong indications that different prong sizes have no influence on essential respiratory parameters or the elimination of pCO2 when using NHF in COPD patients.
More Related Videos
05:39Author Spotlight: A Non-Intubated Video-Assisted Thoracoscopic Surgery with Multimodal Analgesia and Sevoflurane Inhalation Anesthesia
Published on: May 26, 2023
09:31Normothermic Negative Pressure Ventilation Ex Situ Lung Perfusion: Evaluation of Lung Function and Metabolism
Published on: February 14, 2022
Related Concept Videos
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...