ClearLLab 10C reagents panel can be applied to analyze paucicellular samples by flow cytometry
Małgorzata Kajstura1, Tia LaBarge1, Andrew G Evans1
1Department of Pathology and Laboratory Medicine, University of Rochester Medical Center, Rochester, New York, USA.
Cytometry. Part B, Clinical Cytometry
|November 19, 2024
Summary
The ClearLLab 10C Reagents Panel effectively diagnoses leukemias and lymphomas in low cell concentration samples. This flow cytometry method maintains accuracy even with 64-fold lower cell counts, expanding its clinical use.
Area of Science:
- Hematology
- Immunophenotyping
- Clinical Diagnostics
Background:
- The FDA-approved ClearLLab 10C Reagents Panel simplifies leukemia and lymphoma diagnosis via flow cytometry.
- A limitation is its requirement for 3 × 106 cells/mL, which is unachievable for paucicellular samples.
Purpose of the Study:
- To evaluate the reliability of the ClearLLab 10C 10-color panel for analyzing specimens with low cell concentrations.
- To determine if the panel can be effectively used with cell concentrations significantly lower than the FDA-approved standard.
Main Methods:
- Serial dilutions of 16 samples (5 normal, 11 abnormal) were prepared, ranging from 3.0 × 106 to 0.0469 × 106 cells/mL.
- Samples were stained using the B-cell and T-cell panels of the ClearLLab 10C system.
- Mean fluorescence intensity (MFI) was measured, and deviations from the standard concentration were calculated and analyzed using Bland-Altman, correlation, and regression methods.
Main Results:
- Antigen expression patterns remained consistent across all tested cell concentrations.
- The mean deviation of MFI from the standard concentration never exceeded 10% for any antibody.
- Bland-Altman analysis showed similarity between standard and diluted cell suspension results, with high positive correlation.
- A 100% concordance in antigen expression patterns was observed between the FDA-approved concentration and the lowest cell density tested.
Conclusions:
- The ClearLLab 10C panel reliably determines marker expression for leukemias and lymphomas in paucicellular samples.
- The panel is effective with cell concentrations as low as 0.0469 × 106 cells/mL (64-fold lower than FDA-approved).
- This finding significantly broadens the clinical applicability of the ClearLLab 10C platform for flow cytometry diagnostics.