Imaging of Biological Tissues by Desorption Electrospray Ionization Mass Spectrometry
Published on: July 12, 2013
Improved ion detection sensitivity in mass spectrometry imaging using tapping-mode scanning probe electrospray
Yoichi Otsuka1,2, Maki Okada3, Tomomi Hashidate-Yoshida4
1Department of Physics, Graduate School of Science, Osaka University, Toyonaka, Osaka, Japan. otsuka@phys.sci.osaka-u.ac.jp.
Abstract:
Mass spectrometry imaging (MSI) is a promising analytical method to visualize the distribution of lipids in biological tissues. To clarify the relationship between cellular distribution and lipid types in a tissue, it is crucial to achieve both an improvement in ion detection sensitivity and a reduction in the ionization area. We report methods for improving the efficiency of ion transfer to a mass spectrometer and miniaturizing the extraction area of a sample for tapping-mode scanning probe electrospray ionization (t-SPESI), atmospheric pressure sampling, and ionization methods. To verify the efficacy of the new t-SPESI measurement system, MSI was performed on mouse testes with a pixel size of 5 µm. Lipid images of the testes from wild-type (WT) and lysophospholipid acyltransferase 3 (LPLAT3) knockout mice revealed the characteristic distribution of docosahexaenoic acid-containing phospholipids (DHA-PLs). A comparison of the ion images obtained by MSI and optical images of the same tissues stained with hematoxylin and eosin suggested that the distribution of DHA-PLs was significantly altered by spermatogenesis in the WT mouse testes.