Quantitative analysis of patient motion in walk-through PET scanner and standard axial field of view pet scanner
Rabia Aziz1, Jens Maebe2, Florence Marie Muller2,3
1Department of Electronics and Information Systems, Medical Image and Signal Processing, Ghent University, C. Heymanslaan 10, Ghent, Belgium. rabia.aziz@ugent.be.
Background:
Long-axial field-of-view (LAFOV) Positron Emission Tomography (PET) scanners provide high sensitivity, but throughput is limited because of time-consuming patient positioning. To enhance throughput, a novel Walk-Through PET (WT-PET) scanner has been developed, allowing patients to stand upright, supported by an adjustable headrest and hand supports. This study evaluates the degree of motion in the WT-PET system and compares it with the standard PET-CT.
Methods:
Three studies were conducted with healthy volunteers to estimate motion. The first two studies assessed motion in the WT-PET's Design I (Study 1) and Design II (Study 2), while the third study compared motion on a standard PET-CT scanner bed (Study 3). Infrared markers placed on the head, shoulders, chest, and abdomen were tracked and processed using image-processing techniques involving thresholding and connected component analysis. Videos were recorded for normal breathing and breath-holding conditions, and 2D centroids were transformed into 3D coordinates using depth information.
Results:
The results shows a significant reduction in motion during breath-holding, especially for the abdomen. Mean motion distances decreased from 2.63 mm to 2.18 mm in Study 1 and from 2.42 mm to 1.67 mm in Study 2. Statistical analysis revealed notable differences in motion between the WT-PET and mCT scanners. The Shapiro-Wilk test indicated non-normal motion distributions in the head, right shoulder, and abdomen for both systems, leading to the use of the Wilcoxon signed-rank test for all markers. Significant differences were found in the right shoulder (p = 0.0266), left shoulder (p = 0.0004) and chest (p < 0.0001) but no significant differences were observed in the head (p = 0.1327) and abdomen (p = 0.8404).
Conclusion:
This study provides a comprehensive analysis of patient motion in a WT-PET scanner with respect to the standard PET. The findings highlight a significant increase in shoulder and chest motion, while the head and abdomen regions showed more stability.
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET