UV-Vis Spectroscopic Characterization of Nanomaterials in Aqueous Media
Published on: October 25, 2021
Forward-Scattering and Multiple-Scattering Sources of Errors in UV-Visible Spectroscopy of Microspheres
Azizeh Alidoust Ghatar1, Baptiste Auguié1, Eric C Le Ru1
1The MacDiarmid Institute for Advanced Materials and Nanotechnology, School of Chemical and Physical Sciences, Victoria University of Wellington, P.O. Box 600, Wellington 6140, New Zealand.
Abstract:
Conventional UV-visible spectroscopy instruments measure the extinction spectrum of solutions in a transmission configuration. Because of the finite (nonzero) acceptance angle in detection, errors due to forward scattering and multiple scattering can be introduced when measuring scattering samples. We here experimentally quantify these errors using polystyrene spheres of different sizes for two representative analytical/research UV-visible instruments, one based on a single-beam diode array and the other on a double-beam scanning configuration. The measured spectra for particles larger than 1 μm are shown to differ between the two instruments, even at low concentrations, and also vary with concentration (in contradiction with the Beer-Lambert law). We show that systematic errors in the range of 10-40% are common in such measurements. We propose a model accounting for both forward- and multiple-scattering errors and demonstrate its agreement with our experimental results. This model could reduce systematic errors in measurements of scattering samples by up to 40%.
Related Concept Videos
UV–Vis Spectrometers
Ultraviolet and Visible (UV–Vis) Spectroscopy: Overview
UV–Vis Spectroscopy: Molecular Electronic Transitions
UV–Vis Spectrum
The UV–Vis spectrum of a molecule is the plot of its absorbance versus wavelength. The plot is drawn by taking molar...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
NMR Spectrometers: Resolution and Error Correction