Failure of the Jonas prosthesis.

E A Tawil, J G Gregory

    The Journal of Urology
    |April 1, 1986
    PubMed
    Summary

    The Jonas silicone-silver penile prosthesis can fracture, leading to device failure and rigidity loss. X-rays are unreliable for diagnosing these fractures, necessitating clinical evaluation.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Sympathetic innervation and chemical sympathectomy of canine bladder.

    Urology·2011
    Same author

    Treatment of post-orchiectomy hot flashes with transdermal administration of clonidine.

    The Journal of urology·1990
    Same author

    Transrectal ultrasound in stage A1 prostate carcinoma.

    Urology·1989
    Same author

    Impotence. An historic perspective.

    ASAIO transactions·1988
    Same author

    Approaches to insomnia. Helping patients, avoiding pitfalls, doing no harm.

    North Carolina medical journal·1988
    Same author

    Surgical implantation of the artificial urinary sphincter.

    ASAIO transactions·1988

    Area of Science:

    • Urology
    • Biomedical Engineering
    • Medical Device Technology

    Background:

    • The Jonas silicone-silver penile prosthesis is a medical device used for treating erectile dysfunction.
    • Device integrity is crucial for patient outcomes and satisfaction.

    Purpose of the Study:

    • To evaluate the incidence and diagnosis of wire fractures in the Jonas silicone-silver penile prosthesis.
    • To assess the reliability of radiographic imaging in detecting these fractures.

    Main Methods:

    • A review of 27 patients who received the Jonas silicone-silver penile prosthesis.
    • Clinical evaluation including penile rigidity assessment and physical examination.
    • Radiographic imaging (X-ray) for suspected fractures.
    • Surgical exploration and device explantation for definitive diagnosis.

    Main Results:

    • Four previously reported cases of prosthesis fracture were identified.
    • Of the remaining 23 patients, 7 showed signs of fracture (decreased rigidity, audible/palpable crackling, or X-ray evidence).
    • Two patients with suspicious X-rays and one with negative X-rays but reduced rigidity underwent explantation, revealing multiple wire fractures.
    • Radiographs of fractured prostheses demonstrated limited diagnostic accuracy.

    Conclusions:

    • Wire fractures are a complication associated with the Jonas silicone-silver penile prosthesis.
    • Clinical assessment of penile rigidity and physical findings are important indicators of prosthesis failure.
    • Radiography has limited reliability in diagnosing wire fractures of this penile prosthesis.