Residual Stress Homogenization of Hybrid Implants
Marta Sanjuán Álvarez1, Daniel Robles2, Javier Gil Mur1
1Bioengineering Institute of Technology, International University of Catalonia, C/de la Inmaculada 22, Sarrià-Sant Gervasi, 08017 Barcelona, Spain.
Objectives:
Hybrid implants commonly exhibit decreased corrosion resistance and fatigue due to differences in compressive residual stresses between the smooth and rough surfaces. The main objective of this study was to investigate the influence of an annealing heat treatment to reduce the residual stresses in hybrid implants.
Methodology:
Commercially pure titanium (CpTi) bars were heat-treated at 800 °C and different annealing times. Optical microscopy was used to analyze the resulting grain growth kinetics. Diffractometry was used to measure residual stress after heat treatment, corrosion resistance by open circuit potential (EOCP), corrosion potentials (ECORR), and corrosion currents (ICORR) of heat-treated samples, as well as fatigue behavior by creep testing. The von Mises distribution and the resulting microstrains in heat-treated hybrid implants and in cortical and trabecular bone were assessed by finite element analysis. The results of treated hybrid implants were compared to those of untreated hybrid implants and hybrid implants with a rough surface (shot-blasted).
Results:
The proposed heat treatment (800 °C for 30 min, followed by quenching in water at 20 °C) could successfully homogenize the residual stress difference between the two surfaces of the hybrid implant (-20.2 MPa). It provides better fatigue behavior and corrosion resistance (p ˂ 0.05, ANOVA). Stress distribution was significantly improved in the trabecular bone. Heat-treated hybrid implants performed worse than implants with a rough surface.
Clinical Significance:
Annealing heat treatment can be used to improve the mechanical properties and corrosion resistance of hybrid surface implants by homogenizing residual stresses.
More Related Videos
12:19Multi-Scale Modification of Metallic Implants With Pore Gradients, Polyelectrolytes and Their Indirect Monitoring In vivo
Published on: July 1, 2013
07:09In situ Compressive Loading and Correlative Noninvasive Imaging of the Bone-periodontal Ligament-tooth Fibrous Joint
Published on: March 7, 2014
Related Concept Videos
Residual Stresses
Residual Stresses in Circular Shafts
Residual Stresses in Bending
Transformation of Plane Stress
Stress: General Loading Conditions
The shearing force, possessing potential directionality within the plane of the section, is simplified into two component forces running parallel to the x and y axes....
Generalized Hooke's Law