Daily Diagnostic Quality Computed Tomography-on-Rails (CTOR) Image Guidance for Abdominal Stereotactic Body Radiation
Rachael M Martin-Paulpeter1, P James Jensen1, Luis A Perles1
1Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, TX 77030, USA.
Background/Objectives:
Stereotactic body radiation therapy (SBRT) for abdominal targets faces a variety of challenges, including motion caused by the respiration and digestion and a relatively poor level of contrast between the tumor and the surrounding tissues. Breath-hold treatments with computed tomography-on-rails (CTOR) image guidance is one way of addressing these challenges, allowing for both the tumor and normal tissues to be well-visualized. Using isodose lines (IDLs) from CT simulations as a guide, the anatomical information can be used to shift the alignment or trigger a replan, such that normal tissues receive acceptable doses of radiation.
Methods:
This study aims to describe the workflow involved when using CTOR for pancreas and liver SBRT and demonstrates its effectiveness through several case studies.
Results:
In these case studies, using the anatomical information gained through diagnostic-quality CT guidance to make slight adjustments to the alignment, resulted in reductions in the maximum dose to the stomach.
Conclusions:
High-quality imaging, such as CTOR, and the use of IDLs to estimate the doses to OARs, enable the safe delivery of SBRT, without the added complexity and resource commitment required by daily online adaptive planning.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...