Performing Intracochlear Electrocochleography During Cochlear Implantation
Published on: March 8, 2022
A Novel Algorithm to Analyze Multi-Frequency Electrocochleography Measurements to Monitor Electrode Placement During
Eric E Babajanian1, Kanthaiah Koka2, Aniket A Saoji1
1Department of Otolaryngology-Head and Neck Surgery, Mayo Clinic, Rochester, MN 55902, USA.
Objectives:
During cochlear implant (CI) electrode placement, single low-frequency (e.g., 500 Hz) cochlear microphonics (CM) measurements are used to monitor hair-cell function and provide feedback to avert insertion trauma. However, it can be difficult to differentiate between trauma and the electrode's progression through the cochlea when monitored with a single frequency. Multi-frequency CM measurements, while more complex to analyze, can provide more accurate feedback by measuring CM from various locations along the basilar membrane.
Methods:
A new algorithm was developed to analyze multi-frequency CM tracings by comparing amplitude and phase changes across different test frequencies. The new algorithm was evaluated as to its ability to identify drop-alarm instances with the multi-frequency approach, as compared to single-frequency 500 Hz tracings.
Results:
The algorithm presented in this manuscript uses the relationship between CM amplitude and phase changes across frequencies to provide real-time feedback during CI electrode placement. The results show that multi-frequency CM tracings raised an alarm only 0.5 times, as compared to 2.8 instances of alarm raised for the single-frequency 500 Hz CM measurements.
Conclusions:
Multi-frequency CM tracings can help reduce the number of alarms which may be false positives prompting unnecessary electrode manipulations, thereby minimizing the risk of insertion trauma.