Related Experiment Video
Updated: Jun 14, 2026

06:58
A Structured Rehabilitation Protocol for Improved Multifunctional Prosthetic Control: A Case Study
Published on: November 6, 2015
9.4K
Replication of clinical prosthetic sockets for research purposes
Alix Chadwell1,2,3, Laurence Kenney2, Michael Prince2
1School of Healthcare Enterprise and Innovation, Faculty of Medicine, University of Southampton, Southampton, UK.
Prosthetics and Orthotics International
|November 27, 2024
Summary
Researchers developed a new method to accurately clone clinical prosthetic sockets without needing a prosthetist. This technique aids in prosthetic technology research by enabling precise socket replication for improved testing.
Area of Science:
- Biomedical Engineering
- Prosthetics and Orthotics
- Rehabilitation Technology
Background:
- Clinical research in prosthetics often requires standard prosthetic sockets for realistic outcomes.
- A global shortage of prosthetists limits the inclusion of clinical sockets in research studies.
- Engineering-focused labs often conduct studies with anatomically intact individuals, lacking clinical relevance.
Purpose of the Study:
- To present a novel, prosthetist-independent method for accurately cloning clinical standard prosthetic sockets.
- To facilitate more representative and accurate research into prosthetic technologies.
- To enable further research into socket design and electrode placement by removing confounding factors.
Main Methods:
- A silicone-based technique was employed to capture the precise shape of a clinical prosthetic socket.
- The captured shape was converted into a plaster mold for manufacturing identical sockets.
- 3D scanning and CloudCompare software were used to validate the accuracy of the cloned socket shape against the original.
Main Results:
- The proposed technique successfully created an identical clone of the clinical standard prosthetic socket.
- 3D scan analysis showed a mean distance of only 0.16 mm (SD 0.38 mm) between the original and cloned socket meshes.
- The cloning process was demonstrated to be feasible, accurate, and non-destructive.
Conclusions:
- The developed method provides an accurate and accessible way to replicate clinical prosthetic sockets for research.
- This technique overcomes the limitations imposed by the prosthetist shortage, enhancing prosthetic technology assessment.
- The non-destructive nature of the process opens new avenues for research in prosthetic socket design and functionality.