Concordance of grip strength measurement between a digital handheld dynamometer and a Hydraulic dynamometer
Amor A Segura-Duarte1, Gerardo Méndez-Suárez2, Laura P Burgos-Morelos1
1Geriatrics Department.
Background:
Assessment of muscle strength is of great interest for the study of age-associated pathologies, including frailty and sarcopenia. However, the cost of some of the equipment necessary for its evaluation could be prohibitive for its generalized use in some settings.
Objective:
The purpose of the present study was to compare two handheld dynamometers to evaluate the measure of concordance between the measurements. Material and methods The study was conducted in the outpatient Geriatrics clinics where 120 community-dwelling participants aged 60 years and more were evaluated using two devices: the JAMAR® Hydraulic Hand Dynamometer and the digital hand dynamometer CAMRY® EH101.
Results:
We observed a positive correlation between the two devices for dominant hand (DH) Pearson r 0.974 (p < 0.001) and a Spearman ρ 0.973 (p < 0.001) for non-dominant hand (NDH) Lin's coefficient for the DH was 0.98 (p < 0.001) and for the NDH 0.97 (p < 0.001). The intraclass correlation coefficient was 0.971 (95% CI, 0.95-0.97; p < 0.001) for DH and 0.975 (95% CI, 0.96-0.98; p < 0.001) for NDH.
Conclusions:
In this study, the Camry dynamometer was observed to be reliable for measuring hand grip strength in community-dwelling people 60 years or older.
Related Concept Videos
Non-destructive Tests for Concrete Strength
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)