Related Experiment Video
Updated: Jun 6, 2025

Using an Ingestible Telemetric Temperature Pill to Assess Gastrointestinal Temperature During Exercise
Published on: October 7, 2015
Accuracy and precision of non-invasive thermometers compared with the pulmonary artery temperature: a cross-sectional
Rafael Lima Rodrigues de Carvalho1, Mariana Avendanha Victoriano2, Camila Claúdia Campos3
1Adjunct Professor, Nursing School, Universidade Federal da Bahia (UFBA), Salvador (BA), Brazil.
Background:
Temperature fluctuations are critical indicators of a patient's condition in intensive care units (ICUs). While invasive methods offer a more reliable measurement of core temperature, they carry greater risks of complications, limiting their use in most situations. This underscores the need for research evaluating the reliability of non-invasive temperature monitoring methods.
Objectives:
This study aimed to assess the accuracy and precision of four non-invasive temperature measurement techniques compared to pulmonary artery temperature, considered the gold standard.
Design And Setting:
We conducted a cross-sectional clinical study with repeated measures in the ICUs at Hospital das Clínicas da Universidade Federal de Minas Gerais and Hospital Felício Rocho, Belo Horizonte, Brazil.
Methods:
All patients admitted with a pulmonary artery catheter were included. We simultaneously recorded temperatures from the pulmonary artery, axillary area, oral cavity, temporal artery, and tympanic membrane. Bland-Altman plots were employed to assess the agreement between the different temperature measurements.
Results:
A total of 48 patients participated, with a mean age of 54 years. Females comprised 66.67% of the sample. Compared to pulmonary artery temperature, the accuracy and precision (mean and standard deviation) of the non-invasive methods were: axillary (-0.42°C, 0.59°C), oral (-0.30°C, 0.37°C), tympanic membrane (-0.21°C, 0.44°C), and temporal artery (-0.25°C, 0.61°C). Notably, in patients with abnormal body temperature (non-normothermic), only oral and tympanic membrane methods maintained their accuracy and precision.
Conclusions:
The non-invasive thermometers evaluated in this study demonstrated acceptable accuracy and precision (within the clinically relevant threshold of 0.5°C) compared to pulmonary artery temperature. Among the non-invasive methods, the tympanic membrane measurement proved to be the most reliable, followed by the oral method.
More Related Videos
04:34Mouse Body Temperature Measurement Using Infrared Thermometer During Passive Systemic Anaphylaxis and Food Allergy Evaluation
Published on: September 14, 2018
07:20Lavage-induced Surfactant Depletion in Pigs As a Model of the Acute Respiratory Distress Syndrome ARDS
Published on: September 7, 2016
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...