Using a Cyclic Ion Mobility Spectrometer for Tandem Ion Mobility Experiments
Published on: January 20, 2022
Development of a Platform for High-Resolution Ion Mobility Separations Coupled with Messenger Tagging Infrared
Christopher P Harrilal1, Sandilya V B Garimella1, Randolph V Norheim1
1Biological Sciences Division, Pacific Northwest National Laboratory, P.O. Box 999, Richland, Washington 99354, United States.
Abstract:
The ability to uniquely identify a compound requires highly precise and orthogonal measurements. Here we describe a newly developed analytical platform that integrates high resolution ion mobility and cryogenic vibrational ion spectroscopy for high-precision structural characterizations. This platform allows for the temporal separation of isomeric/isobaric ions and provides a highly sensitive description of the ion's adopted geometry in the gas phase. The combination of these orthogonal structural measurements yields precise descriptors that can be used to resolve between and confidently identify highly similar ions. The unique benefits of our instrument, which integrates a structures for lossless ion manipulations ion mobility (SLIM IM) device with messenger tagging infrared spectroscopy, include the ability to perform high-resolution ion mobility separations and to record the IR spectra of all ions simultaneously. The SLIM IM device, with its 13 m separation path length, allows for multipass experiments to be performed for increased resolution as needed. It is integrated with an Agilent qTOF MS where the collision cell was replaced with a cryogenically held (30 K) TW-SLIM module. The cryo-SLIM is operated in a novel manner that allows ions to be streamed through the device and collisionally cooled to a temperature where they can form noncovalently bound N2 complexes that are maintained as they exit the device and are detected by the TOF mass analyzer. The instrument can be operated in two modes: IMS+IR where the IR spectra for mobility-selected ions can be recorded and IR-only mode where the IR spectra for all mass-resolved ions can be recorded. In IR-only mode, IR spectra (400 cm-1 spectral range) can be recorded in as short as 2 s for high throughput measurements. This work details the construction of the instrument and modes of operation. It provides initial benchmarking of CCS and IR measurements to demonstrate the utility of this instrument for targeted and untargeted approaches.
More Related Videos
07:33Analyzing Protein Architectures and Protein-Ligand Complexes by Integrative Structural Mass Spectrometry
Published on: October 15, 2018
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
Related Concept Videos
High-Resolution Mass Spectrometry (HRMS)
Ion-Exchange Chromatography
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Matrix-Assisted Laser Desorption Ionization (MALDI)
The analyte of interest, a biomolecule or a mixture of biomolecules, is mixed with a suitable matrix material. The...
Mass Spectrometers
Mass Analyzers: Common Types