Related Experiment Video
Updated: Jun 6, 2025

Reductive Electropolymerization of a Vinyl-containing Poly-pyridyl Complex on Glassy Carbon and Fluorine-doped Tin Oxide Electrodes
Published on: January 30, 2015
Highly sensitive voltammetric determination of the fungicide fenhexamid using a cost-effective and disposable pencil
Teslime Erşan1, Didem Giray Dilgin2, Ayhan Oral1
1Department of Chemistry, Faculty of Science, Çanakkale Onsekiz Mart University, Çanakkale, 17020, Türkiye.
Abstract:
A differential pulse voltammetric (DPV) method is proposed for the highly sensitive determination of fenhexamid (FHX) based on both electrooxidation and electroreduction processes using a disposable and cost-effective pencil graphite electrode (PGE). The electrochemical oxidation and reduction mechanisms of FHX at the PGE were elucidated by recording cyclic voltammograms at various pH values of Britton-Robinson buffer (BRB) solutions at a scan rate of 50 mV s-1 and different scan rate values in the range 10-400 mV s-1 at selected pH of BRB (pH 2.0). Differential pulse voltammograms recorded under optimized conditions revealed an oxidation peak of FHX around + 0.65 V and a reduction peak of FHX around + 0.45 V. The DPV analysis of FHX revealed two linear ranges: 0.001-0.01 µmol L-1 and 0.01-5.0 µmol L-1 for the anodic peak, and 0.001-0.1 µmol L-1 and 0.1-5.0 µmol L-1 for the cathodic peak. The limits of detection were 0.34 nmol L-1 and 0.32 nmol L-1 for the anodic and cathodic peaks, respectively. The proposed methodology demonstrated satisfactory selectivity, as selected pesticides, certain electroactive compounds, and cationic species tested did not interfere with the voltammetric determination of FHX, particularly during its reduction. The recovery results, showing values close to 100% obtained from the analysis of real samples spiked with FHX, indicated that this methodology can accurately determine FHX in water and soil samples.
Related Concept Videos
Voltammetric Techniques: Linear-Scan (E vs Time)
Voltammetry: Overview
A voltammetric cell uses three electrodes: a working electrode, a reference electrode, and an auxiliary electrode. The redox reactions occur in the working...
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Voltammetric Techniques: Pulse Voltammetry
Voltammetric Techniques: Cyclic Voltammetry