Use of a modified Foley catheter in patients with Montgomery T-tubes
1Department of Otolaryngology, Faculty of Medicine, Universiti Malaya, Lembah Pantai, Kuala Lumpur, 50603, Malaysia.
Background:
The T-tube poses significant challenges for positive pressure ventilation.
Method:
We describe the modification of a Foley catheter by trimming off the proximal urine drainage port, and using this opening to pass a metal stylet within the catheter, providing additional rigidity and allowing easier manipulation of the catheter tip. The catheter tip is then positioned in the superior limb of the T-tube, and the balloon inflated under direct visualization to occlude the opening and allow positive pressure ventilation through the external limb of the T-tube.
Conclusion:
This simple yet effective technique can be considered in patients with T-tubes.
Related Concept Videos
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Tracheostomy Suctioning II: Procedure
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...