Comparison of pulse pressure and stroke volume variations measured by three monitors in high-risk surgical patients
Barbora Cenková1,2,3, Miloš Chobola2, Vladimír Šrámek2,3
1International Clinical Research Center, St. Anne's University Hospital Brno, Brno, Czech Republic.
Introduction:
Dynamic indices of fluid responsiveness (FR) such as pulse pressure variation (PPV) and stroke volume variation (SVV) differ among hemodynamic monitors, which use proprietary algorithms, and vary even over a short period of time. We aimed to compare the baseline values, fluctuation and predictive value for FR of PPV and SVV measured by three minimally invasive monitors.
Patients And Methods:
Twenty patients undergoing high-risk abdominal surgery were included and 45 fluid challenges were analysed. The patients were simultaneously monitored using Carescape B650, LiDCO Rapid and FloTrac/Vigileo system. Cardiac output (CO), PPV and SVV were recorded before and after the fluid challenge of 500 ml of balanced crystalloid solution. An increase in CO ≥ 15 % defined fluid responders. Concurrently recorded arterial waveform was used for offline calculation of PPV.
Results:
Mean baseline values of the indices ranged between 8.6 % and 13.4 %. LiDCO showed higher fluctuation of indices compared to the other monitors. Area under the receiver operating characteristic curve (AUROC) ranged from 0.71 to 0.76 with optimal cut-off value between 7.5 % and 13.9 %. AUROC increased for all indices when FR was defined as an increase in stroke volume. Furthermore, a decrease in PPV or SVV after fluid challenge (ΔPPV, ΔSVV) proved a better marker of FR (AUROC 0.82-0.93) than baseline values with a uniform threshold of approximately -3%.
Conclusions:
Although a significant range of baselines variations and optimal cut-off values was observed, the predictive value of PPV and SVV from all the monitors was only moderate and comparable. Nevertheless, ΔPPV and ΔSVV appear to be a reliable and device-independent markers of FR.
More Related Videos
06:04Author Spotlight: Assessing the Cardiovascular Profile of Patients with Metabolic Syndrome
Published on: September 27, 2024
08:51Multiple Intravenous Bolus Dosing and Invasive Hemodynamic Assessment in a Hypoxia-Induced Mouse Pulmonary Artery Hypertension Model
Published on: November 11, 2022
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure