A commissioning protocol for portal imaging-based radiotherapy in vivo dosimetry systems
Marco Esposito1,2, Riccardo Baldoni3, Evy Bossuyt4
1Medical Physic, The Abdus Salam International Centre for Theoretical Physics, Trieste, Italy.
Background And Purpose:
With the availability of commercial electronic portal imaging detector-based in vivo dosimetry (EPID-based IVD) solutions, many radiotherapy departments are adopting this technology. However, comprehensive commissioning guidance is lacking. This study aims to provide a protocol for testing the accuracy and sensitivity of EPID-based IVD systems.
Material And Methods:
The protocol was tested across four institutions using two different systems. Accuracy was evaluated with homogeneous slab phantoms using different square regular fields, and clinical plans in a CIRS lung phantom. Multiple forward and back-projected algorithm implementations were assessed for different energies. Sensitivity analysis in the lung phantom examined responses to setup errors, anatomical variations, and delivery errors.
Results:
In homogeneous phantoms, over 85 % of pixels passed the 5 %/2mm gamma criteria, except for the 2x2 cm2 field. In the lung phantom, all systems and implementations achieved over 95 %-pixel pass rates at the 2 %/2mm criterion for volumetric modulated arc therapy (VMAT) plans. For conformal radiation therapy (3DCRT) plans, one system implementation showed poor accuracy, with over 90 % agreement only at the 5 %/2mm criterion. Considering all systems and implementations, average sensitivity and specificity for CRT plans ranged from 0.92 and 0.42 (at 2 %/2mm) to 0.71 and 0.52 (at 5 %/2mm), while for VMAT plans ranged from 0.41 and 0.81 (at 2 %2mm) to 0.37 and 0.81 (at 5 %/2mm).
Conclusion:
We successfully developed a protocol to commission EPID IDV systems. It was found that not all systems and implementations achieved satisfactory accuracy and sensitivity, emphasising the need for thorough commissioning and benchmarking.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET