Empirical correction decomposition method (ECDM): enhancing accuracy of quantitative measurement in spectral CT
Chengmin Wang1,2, Zhe Wang1,3, Mohan Li1,3
1Beijing Engineering Research Center of Radiographic Techniques and Equipment, Institute of High Energy Physics, Chinese Academy of Sciences, Beijing 100049, People's Republic of China.
Abstract:
Objective. Spectral CT and material decomposition methods are crucial for precise material identification and quantitative composition analysis in preclinical research and clinical diagnosis. The empirical material decomposition method is widely used for its straightforward modeling approach, independence from spectral and detector response knowledge, and operational convenience. However, this method has limited decomposition accuracy and its precision depends on the choice of calibration phantoms.Approach. To address these issues, we propose an empirical correction decomposition method (ECDM). The innovation of this method lies in its ability to conveniently estimate and correct empirical decomposition errors using a specially designed calibration phantom. First, the specially designed calibration phantom for ECDM undergoes empirical decomposition initially to establish the relationship between decomposition errors and decomposition values. Then, ECDM estimates and corrects the error of empirical decomposition values.Main results. In the phantom experiments, ECDM improves the decomposition accuracy of empirical methods, effectively reducing the different decomposition errors caused by four different sizes of calibration phantoms from a maximum of 144% to within 25%. In the mouse experiments, ECDM achieves accurate quantification of contrast agents in biological tissues, outperforming the other two methods. The absolute error percentages of ECDM in the decomposition results of the two standard iodine solutions are both less than 5%.Significance. ECDM significantly improves decomposition accuracy and reduces the impact of the size of the empirical calibration phantom. Overall, our method based on spectral CT is very convenient and practical for the quantitative measurement in biomedical applications.
More Related Videos
10:42In Depth Analyses of LEDs by a Combination of X-ray Computed Tomography CT and Light Microscopy LM Correlated with Scanning Electron Microscopy SEM
Published on: June 16, 2016
07:05Applying Hyperspectral Reflectance Imaging to Investigate the Palettes and the Techniques of Painters
Published on: June 18, 2021
Related Concept Videos
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
Distance Corrections
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
NMR Spectrometers: Resolution and Error Correction