Measurement of the AH and HV interval and His bundle duration: interobserver and intraobserver variation

Insights

Intracardiac electrophysiology testing reliability varies. AH and HV intervals show moderate reliability for clinical decisions, but His bundle duration measurements are unreliable.

Area of Science:

  • Electrophysiology
  • Cardiology
  • Medical Measurement

Background:

  • Conduction system intervals are crucial for clinical decisions during electrophysiologic testing.
  • Variability in measurements can impact patient management.
  • Assessing the reliability of these measurements is essential.

Purpose of the Study:

  • To evaluate the interobserver and intraobserver reliability of AH interval, HV interval, and His bundle duration measurements.
  • To determine if these measurements are reliable enough for clinical decision-making.

Main Methods:

  • Fifty patients undergoing intracardiac electrophysiologic testing were included.
  • Three investigators measured AH interval, HV interval, and His bundle duration.
  • The square of the correlation coefficient (r2) was used to assess measurement reliability.

Main Results:

  • AH interval interobserver r2 ranged from 0.55-0.68; intraobserver r2 ranged from 0.60-0.88.
  • HV interval interobserver r2 ranged from 0.42-0.63; intraobserver r2 ranged from 0.489-0.73.
  • His bundle duration showed lower reliability (interobserver r2: 0.42-0.52; intraobserver r2: 0.19-0.52).

Conclusions:

  • AH and HV intervals demonstrate sufficient reliability for clinical use alongside patient history and ECG.
  • His bundle duration measurements lack reliability with standard techniques and should not guide clinical decisions.