Quantitation of biological retinoids by high-pressure liquid chromatography: primary internal standardization using
Abstract:
A single method is described for quantitation of 14 retinoids found in biological material. The method consists of reversed-phase HPLC, internal standardization, and carrier extraction procedures with three synthetic retinoids. Primary standardization of HPLC uv detector is achieved using tritiated all-trans-retinoic acid, all-trans-retinol, all-trans-retinyl palmitate, and all-trans-retinyl acetate. Extraction methods are standardized by correlating the uv absorbance of retinoids at 340 nm with radioactivity of tritiated retinoids of known specific activity. Quantitation of 10 pg of tritiated or 5 ng of nonradioactive retinoid per 0.1 g sample in a polarity range from 4-oxo-retinoic acid to retinyl stearate can be achieved in a single, 50-min chromatographic run. A single HPLC pump, a C18 reversed-phase analytical column, a multistep three-solvent gradient, and inexpensive solvents based on methanol, water, and chloroform comprise this cost-effective chromatographic system. Our primary standardization method allows investigators employing different procedures to compare results between laboratories by standardizing the HPLC uv detector with commercially available tritiated retinoids. With this method we were able to quantitate nanomolar amounts of endogenous retinoic acids and retinyl esters, that "HPLC uv only" conditions usually would not detect in the circulation and liver of rats under physiological conditions.