A Structured Rehabilitation Protocol for Improved Multifunctional Prosthetic Control: A Case Study
Published on: November 6, 2015
Real-time adaptive cancellation of TENS feedback artifact on sEMG for prosthesis closed-loop control
Byungwook Lee1, Kyung-Soo Kim1, Younggeol Cho2
1Department of Mechanical Engineering, Mechatronics Systsems and Control, Korea Advanced Institute of Science and Technology, Deajeon, Republic of Korea.
Introduction:
The prosthetic hand has been aimed to restore hand functions by estimating the user's intention via bio-signal and providing sensory feedback. Surface electromyogram (sEMG) is a widely used signal, and transcutaneous electrical nerve stimulation (TENS) is a promising method for sensory feedback. However, TENS currents can transmit through the skin and interfere as noise with the sEMG signals, referred to as "Artifact," which degrades the performance of intention estimation.
Method:
In this paper, we proposed an adaptive artifact removal method that can cancel artifacts separately across different frequencies and pulse widths of TENS. The modified least-mean-square adaptive filter uses the mean of previous artifacts as reference signals, and compensate using prior information of TENS system. Also temporal separation for artifact discrimination is applied to achieve high artifact removal efficiency. Four sEMG signals-two from flexor digitorum superficialis, flexor carpi ulnaris, extensor carpi ulnaris-was collected to validate signals both offline and online experiments.
Results And Discussion:
We validated the filtering performance with twelve participants performing two experiments: artifact cancellation under variable conditions and a real-time hand control simulation called the target reaching experiment (TRE). The result showed that the Signal-to-Noise Ratio (SNR) increased by an average of 10.3dB, and the performance of four TRE indices recovered to the levels similar to those without TENS. The proposed method can significantly improve signal quality via artifact removal in the context of sensory feedback through TENS in prosthetic systems.
More Related Videos
09:14Surface Electromyographic Biofeedback as a Rehabilitation Tool for Patients with Global Brachial Plexus Injury Receiving Bionic Reconstruction
Published on: September 28, 2019
07:28A Method for Evaluating Timeliness and Accuracy of Volitional Motor Responses to Vibrotactile Stimuli
Published on: August 2, 2016
Related Concept Videos
Feedback control systems
Linear feedback systems are theoretical models that simplify analysis and design. These systems operate under the principle that their output is directly proportional to their input within certain ranges. For instance, an amplifier in a control system behaves linearly as long as the input signal remains within a specific range. However, most physical systems exhibit inherent nonlinearity...
Effects of feedback
Feedback significantly modifies the gain of a control system. The gain of a system without feedback is altered by a factor of one plus GH, where G represents...