Computerized QT and QTc Measurements From Bedside ICU Monitors Are Similar to Those Derived From a Standard 12-Lead
Arthur Murray1, Karolina Ho2, Thomas J Hoffmann3
1Benioff Children's Hospital-Oakland, Oakland, California, USA.
None:
QT/QTc prolongation is associated with an increased risk for torsade de pointes. In at-risk hospitalized patients, it is common to obtain a standard 12-lead electrocardiogram (ECG) for this assessment, but this interrupts patient care. Our hospital recently introduced bedside monitors in the intensive care unit (ICU) with continuous QT/QTc software. However, only four of the seven available ECG leads are used.
Purpose:
Evaluate the agreement between computerized QT/QTc measurements from the bedside monitor (four leads) and a time-matched standard 12-lead ECG.
Design:
Prospective observational study in three adult ICUs.
Methods:
QT/QTc measurements were obtained from a convenience sample, and the two ECG types were ≤ 30 min apart. Agreement was evaluated using Bland-Altman analysis.
Results:
A total of 120 patients were evaluated for inclusion, and 60 (50%) had a 12-lead ECG for comparison. The mean bias difference for QT measurements was not statistically different (β = -2.47, 95% CI = 5.50 to -11.05; p = 0.44; limits of agreement (LOA) = -64.37 to 59.44). Similar non-statistical differences were observed for QTc (β = -3.20, 95% CI = 5.50 to -11.05; p = 0.44; LOA = -67.43 to 61.03).
Conclusion:
There was good agreement for both QT and QTc measurements between the two methods. These pilot data are promising and suggest QT/QTc measurements from bedside monitors (four leads) may be an acceptable alternative to obtaining additional standard 12-lead ECGs. Given that half of the ICU patients screened did not have a 12-lead ECG recorded, bedside monitor QT/QTc's could identify at-risk patients. However, an evaluation in a larger sample and non-ICU patients is warranted.
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
Holter Monitor: 24-Hour Monitoring
Imaging Studies for Cardiovascular System V: CT