Utility of the Novel Auto + A3 Threshold Function in Maintaining Atrioventricular Synchrony
Ryo Ohinata1, Naoya Inoue1,2, Shuji Morikawa1,2
1Department of Cardiology, Chutoen General Medical Center, Kakegawa, Shizuoka, Japan.
Introduction:
The VDD-leadless pacemaker aims to maintain high atrioventricular (AV) synchrony, but proper sensing setup is crucial.
Methods And Results:
This case was an elderly woman with an AV block who received a dual-chamber pacemaker. Due to ventricular pacing failure, a Micra AV2 was implanted. However, postimplantation, high sinus rate, and frequent premature ventricular contractions were noted. Therefore, the auto + A3 threshold function was used in the atrial-sensing setup. A few days later, the MAM test confirmed good AV synchrony.
Conclusion:
The auto + A3 threshold function equipped in the Micra AV2 may offer to maintain high AV synchrony.
Related Concept Videos
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Mechanism of Cardiac Arrhythmias
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow...
Electrophysiology of Normal Cardiac Rhythm
Antiarrhythmic Drugs: Class III Agents as Potassium Channel Blockers