Assessment of Cardiac Function and Myocardial Morphology Using Small Animal Look-locker Inversion Recovery SALLI MRI in Rats
Published on: July 19, 2013
Free-Breathing Ungated Radial Simultaneous Multi-Slice Cardiac T1 Mapping
Johnathan V Le1,2, Jason K Mendes1, Konstantinos Sideris3
1Utah Center for Advanced Imaging Research (UCAIR), Department of Radiology and Imaging Sciences, University of Utah, Salt Lake City, Utah, USA.
A new free-breathing, ungated cardiac T1 mapping method (FURST) shows comparable results to standard breath-held MOLLI sequences. This advance offers a viable alternative for patients with arrhythmias like atrial fibrillation (AF).
Area of Science:
- Cardiovascular Magnetic Resonance Imaging
- Cardiac T1 Mapping
- Medical Imaging Technology
Background:
- Standard Modified Look-Locker imaging (MOLLI) T1 mapping requires breath-holding and ECG gating, posing challenges for patients with atrial fibrillation (AF).
- A need exists for free-breathing and ungated cardiac T1 mapping techniques to improve patient accessibility and data acquisition.
Purpose of the Study:
- To develop and evaluate a novel free-breathing, ungated radial simultaneous multi-slice (SMS) cardiac T1 mapping framework, termed FURST.
- To assess the performance and comparability of the FURST framework against conventional MOLLI sequences.
Main Methods:
- A retrospective cohort study involving 17 canine and 7 human subjects was conducted.
- Pre- and post-contrast T1 mapping was performed using both the FURST sequence and conventional MOLLI sequences at 1.5T and 3.0T.
- Analysis included segment-wise comparisons of T1 values and extracellular volume (ECV) using correlation and Bland-Altman plots, alongside reproducibility and image quality assessments (PIQUE).
Main Results:
- The FURST framework demonstrated no significant differences in pre-contrast T1, post-contrast T1, or ECV reproducibility compared to MOLLI.
- Image quality, assessed by PIQUE scores, was also comparable between FURST and MOLLI.
- The mean ECV difference between FURST and MOLLI was minimal ().
Conclusions:
- The developed free-breathing, ungated radial SMS (FURST) framework yields T1 and ECV maps of similar quality and reproducibility to MOLLI.
- FURST represents a promising alternative for cardiac T1 mapping, particularly beneficial for patients unable to perform breath-holds or with arrhythmias.