Related Experiment Video
Updated: Jun 5, 2025

Author Spotlight: Microbial Control and Monitoring Strategies for Cleanroom Environments and Cellular Therapies
Published on: March 17, 2023
Mobile unidirectional airflow device reduces air contamination in an intravitreal outpatient clean room setting
Tito Fiore1, Nicola De Santi2, Donatella Pietrella3
1Department of Medicine Surgical, Section of Ophthalmology, University of Perugia, S. Maria Della Misericordia Hospital, 06156, Perugia, Italy.
Purpose:
To evaluate and compare the mycobacterial load using a mobile laminar airflow (LAF) device in an IVI-dedicated outpatient clean room (OCR) without ventilation systems and in a hospital-based operating theatre (HOT).
Methods:
This case-control study was conducted in 2 different settings: OCR and HOT during a series of intravitreal injections (IVIs). The Air Microbial analysis was performed using a Surface Air System instrument at three different moments during the IVI sessions in both settings: at the operative site (OS), four meters from the OS (DOS) and in the disinfection room (DR).
Results:
At the OS in the OCR, the microbial load was 20.33 ± 16.51 CFU/m3 at T-1; 1.00 ± 1.00 CFU/m3 at T0; 6.67 ± 6.03 CFU/m3 at T1; 10.00 ± 9.54 CFU/m3 at T2 and 4.33 ± 3.06 CFU/m3 at T3. At the OS in the HOT, the CFU were respectively 0.00 ± 0.00 at T0; 0.67 ± 0.58 at T1; 0.00 ± 0.00 at T2 and 0.00 ± 0.00 T3.
Conclusions:
This standardized sterile technique in an OCR setting, with a mobile LAF, reduces the microbial colonies to levels that are safe and comparable to those recorded in the HOT setting, thus potentially improving the safety of IVIs in an OCR setting.
Related Concept Videos
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...