Direct Pressure Monitoring Accurately Predicts Pulmonary Vein Occlusion During Cryoballoon Ablation
Published on: February 26, 2013
Achieving reduced radiation exposure with maintained fluoroscopy effectiveness using ultralow-dose settings in
Takashi Kaneshiro1, Sadahiro Murota1, Takeshi Nehashi1
1Department of Cardiovascular Medicine Fukushima Medical University Fukushima Japan.
Background And Aims:
Optimization of fluoroscopic image quality for reducing radiation exposure in cryoballoon pulmonary vein isolation (CB-PVI) has not yet been fully investigated. Therefore, we tried to compare the radiation doses among three different X-ray system settings.
Methods:
Consecutive 148 patients scheduled for their first CB-PVI were prospectively enrolled: low dose with the use of an anti-scatter grid for the first 51 patients (LD + G group), low dose without an anti-scatter grid for the subsequent 46 patients (LD-G group), and ultralow dose (ULD group) with an anti-scatter grid for the remaining 51 patients. We compared the radiation doses required to complete CB-PVI procedures among the groups. There were 27 patients for whom CB-PVI was performed without cine acquisition, but with fluoroscopy only, and the radiation doses were also compared.
Results:
The median procedure time and fluoroscopy time were 119 and 35.5 min, respectively, with no significant differences among the groups. The median cumulative air Kerma (AK) decreased in both the LD-G group (71.8 mGy, p < .001) and the ULD group (73.0 mGy, p < .001), compared to the LD + G group (145.0 mGy). Among 27 patients who underwent CB-PVI without cine acquisition, the median cumulative AK further decreased in both the LD-G group (31.4 mGy, p < .05) and the ULD group (22.7 mGy, p < .01), compared to the LD + G group (64.6 mGy).
Conclusion:
Using an ULD X-ray setting and avoiding cine acquisition, we can reduce radiation exposure, while ensuring the necessary fluoroscopy time for the CB-PVI procedure.