Quantifying the Volume of Residual Air in Commercial Intravenous Fluids and Assessing the Stability of Airless
Lyle Babcock1,2, Hayley Brawley1,2, Christiaan van Nispen3
1Clinical and Operational Space Medicine Innovation Consortium (COSMIC), 59th Medical Wing, Lackland Air Force Base, TX 78236, USA.
Introduction:
Commercial off-the-shelf (COTS) intravenous fluid (IVF) containers contain residual air, introducing the risk of venous air embolism (VAE). Venous air embolism occurs when air displaces blood flow in vasculature. The danger from residual air is often negligible in terrestrial settings, where gravitational forces generate buoyancy, pushing residual air to the top of the IVF container. However, in microgravity there is no buoyancy to separate liquid and gas layers. We performed experiments to quantify the amount of air in COTS IVF containers (Experiment 1) and identify the variables that affect the stability of sterilely produced airless containers (Experiment 2).
Materials And Methods:
Experiment 1: Residual air was quantified across varying volumes (100, 250, 500, and 1,000 mL), container design, and manufacturer (B. Braun, Baxter, ICU Medical, and Grifols) of 0.9% NaCl COTS IVFs. Each container was assessed for absolute volumes of air, as well as air:fluid ratios normalized to 1,000 mL. Experiment 2: 1,000 mL IVF containers from 3 manufacturers were filled with either (1) 100% saline or (2) 95% saline and 5% air by volume. Containers were stored for 168 days at 25°C or 40 °C. The containers were optically imaged to quantify the accumulation of air within each IVF container.
Results:
Experiment 1: There was a trend toward larger container sizes and greater absolute volumes of residual air (R2 = 0.964). However, the smallest air:volume ratio occurred in the Baxter 500 mL VIAFLO Container (18.9 ± 3.8 mL air; 2.3% air by volume), whereas the largest ratio occurred in the B. Braun 250 mL EXCEL Container (55.0 ± 9.3 mL; 22.0% air by volume). Experiment 2: By day 168, 6 experimental containers had ruptured and 100% of the surviving containers (30/36) had an increase in air as compared to baseline. Containers placed at 40 °C had a larger increase in air (27.7 ± 6.6 mL) compared to containers stored at 25 °C (7.5 ± 4.1 mL; P < .0001).
Conclusions:
Residual air has a wide variety of volumes in COTS IVFs. The average amount of residual air is high enough to contribute to clinically significant VAEs, although unlikely to be fatal. If airless IVF containers are produced for exploration missions, a progressive increase in the amount of residual air should be expected. Extremes of temperatures and humidity will increase the reaccumulation of residual air and decrease the shelf-life of airless IVFs.
Related Concept Videos
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Respiratory Volumes and Capacities
Respiratory Capacities
One key metric is the Inspiratory Capacity (IC), which represents the maximum amount of air that can be inhaled with full effort. IC is calculated by summing the tidal volume and inspiratory reserve volume, typically ranging from 2.4 to 3.6 liters.
The Functional Residual Capacity (FRC) represents the air in the...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Lung Capacity