Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Magnetic field quality conversion factors experimentally measured in clinical MR-linac beams for seven MR-compatible
Nathan Orlando1,2, Jennie Crosby3, Carri Glide-Hurst3,4
1Odette Cancer Centre, Sunnybrook Health Sciences Centre, Toronto, Ontario, Canada.
Purpose:
The purpose of this work was to experimentally quantify MR-compatible ionization chamber response for 1.5T Elekta Unity and 0.35T ViewRay MRIdian MR-linac systems through the determination of the magnetic field quality conversion factor, kB,Q.
Methods:
Seven MR-compatible ionization chamber models from Standard Imaging and PTW were evaluated. Both the quality conversion factor kQ and the magnetic field quality conversion factor kB,Q were experimentally determined through a cross-calibration method. Specifically, the ratio of absorbed dose measured with a reference A1SL chamber under reference conditions to corrected output measured with each test chamber at the same point of measurement allowed for the determination of kB,Q. The angular dependence of the magnetic field quality conversion factor for MR-compatible chamber models was assessed for the 1.5T Elekta Unity system by measuring kB,Q with the chamber axis and magnetic field direction aligned at cardinal angles (0°, 90°, 180°, 270°).
Results:
Beam quality conversion (kQ) factors for MR-compatible ionization chambers measured in a standard linac beam showed an average percent difference of -0.09 ± 0.18% compared to computed kQ values for their conventional chamber versions. Similarly, magnetic field quality conversion (kB,Q) factors for corresponding MR and non-MR ionization chamber models measured using the same cross-calibration technique demonstrated average percent differences of -0.1 ± 0.3% and 0.0 ± 0.2% for the Elekta Unity and ViewRay MRIdian, respectively. Investigation of the angular dependence of this correction factor demonstrated identical chamber response for equivalent MR-compatible and conventional chamber models.
Conclusions:
This work provides critical experimental validation of MR-compatible ionization chamber performance, with a direct comparison of measured kB,Q values to corresponding conventional chamber models demonstrating nearly equivalent chamber response. kB,Q values determined using our experimental method will serve as an important reference for upcoming MR-linac reference dosimetry protocols and ultimately represent an important step towards accurate output calibration of MR-linac systems.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
14:11Quantification of Hydrogen Concentrations in Surface and Interface Layers and Bulk Materials through Depth Profiling with Nuclear Reaction Analysis
Published on: March 29, 2016
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
Magnetic Susceptibility and Permeability
When diamagnetic materials are placed under an external magnetic field, the moments opposite to the field are induced. Hence, the susceptibility for diamagnets has a minimal negative value of 10-5–10-6. Since...
Magnetic Field due to Moving Charges
Consider a point charge moving with a constant velocity. Like the electric field, the magnetic field at any point is directly proportional to the magnitude of the charge and inversely proportional to the square of the distance between the source point and the field point. However, unlike the electric field, the magnetic field is always perpendicular to the plane containing the line...
Magnetic Flux
Suppose a surface is divided into elements of area dA. For each element, the component of the magnetic field that is normal to the...
Magnetic Resonance Imaging
Magnetic Field Due To A Thin Straight Wire