Apparent binding of ionized calcium by various buffers.
Clinical Chemistry
|February 1, 1985
Summary
This study evaluated buffers for measuring ionized calcium (Ca2+). HEPES buffer is recommended for aqueous reference solutions due to its stable pH and linear Ca2+ response across different instruments.
Area of Science:
- Analytical Chemistry
- Biochemistry
- Clinical Chemistry
Background:
- Accurate measurement of ionized calcium (Ca2+) is crucial for various physiological and clinical applications.
- Ionic strength and pH significantly influence Ca2+ measurements, necessitating careful control in reference solutions.
- Selecting appropriate buffers is essential for maintaining stable conditions and ensuring reliable Ca2+ quantification.
Purpose of the Study:
- To investigate the impact of different buffer substances and concentrations on the accuracy of ionized calcium (Ca2+) measurements.
- To identify a suitable buffer for use in aqueous reference solutions for Ca2+ determination.
- To assess the performance of buffers concerning electrode bias and linearity of response.
Main Methods:
- Six different buffer substances were tested in aqueous solutions with controlled ionic strength (160 mmol/L) and varying calcium chloride (CaCl2) concentrations.
- Ionized calcium (Ca2+) levels were measured using ion-selective electrodes.
- The effect of increasing buffer concentrations and solution pH on Ca2+ measurements and electrode bias was analyzed.
- The performance of 4-(2-Hydroxyethyl)-1-piperazineethanesulfonic acid (HEPES) was specifically evaluated for pH stability and linearity of Ca2+ response.
Main Results:
- Measured Ca2+ levels decreased with increasing ionic strength and pH.
- Tris buffer exhibited a positive Ca2+ electrode bias, while the other five buffers showed a negative bias with increasing concentration.
- HEPES buffer at 10 mmol/L effectively maintained a stable pH (7.39) in reference solutions.
- HEPES demonstrated a linear Ca2+ response across instruments from three different suppliers.
Conclusions:
- Ionic strength and pH are critical factors affecting Ca2+ electrode measurements.
- HEPES buffer is a suitable choice for aqueous reference solutions due to its ability to maintain stable pH and provide a linear Ca2+ response.
- The findings provide guidance for selecting buffers to improve the accuracy and reliability of ionized calcium measurements.