Related Experiment Video
Updated: Jun 5, 2025

Using Q Suture to Enhance Resistance to Gap Formation and Tensile Strength of Repaired Flexor Tendons
Published on: June 3, 2020
How many knots are necessary to achieve knot security of two high strength suture tapes? A biomechanical comparative
Tatjana Pastor1,2, Ivan Zderic1, Kenneth P van Knegsel1,3
1AO Research Institute Davos, Davos, Switzerland.
Introduction:
According to current clinical practice, a minimum of 7 knots are required to provide secure hold in high-strength sutures. A new technology featuring a suture tape with a salt-infused silicon core has been recently developed, potentially reducing the number of needed knots.
Aims:
to (1) assess the influence of number of knots on tape security, (2) evaluate the effect of different ambient conditions on knot security, and (3) compare the biomechanical competence of the novel versus a conventional suture tape.
Materials And Methods:
A conventional suture tape (ST, SutureTapetm) was considered for knot tying together with the novel suture type (DT, Dynatapetm). Specimens were assigned to receive different number of knots, ranging from 3 to 7, and to be exposed to different media during tying-air (dry), saline solution (wet), and fat (fatty-wet). Seven specimens were considered for each suture type, knot number and ambient condition. With knotted sutures mounted between two roller bearings, quasi-static tensile ramp tests were performed to evaluate knot slippage, ultimate force at rupture, and minimum number of knots preventing suture unraveling for each suture tape and condition.
Results:
Whereas the ST ruptured without unraveling with a minimum of 6 knots in all specimens and ambient conditions, the minimum number of knots for a DT rupture without unraveling was 6 in dry, 4 in wet, and 5 in fatty-wet condition. Ultimate force at rupture with a minimum number of needed knots did not differ significantly between ST and DT (p ≥ 0.067), in contrast to knot slippage that was significantly bigger for ST versus DT in wet and fatty-wet conditions(p ≤ 0.001).
Conclusions:
In fatty-wet conditions-related to open surgery-the novel Dynatapetm suture tape requires 5 instead of 7 knots to achieve their security. In wet conditions-related to arthroscopic surgery-this number can be reduced to 4 knots. In contrast, the conventional SutureTapetm needs 6 knots to provide security in all conditions.
Related Concept Videos
Cable: Problem Solving
Bending of Members Made of Several Materials
Hooke's Law determines stress in each material, stating that stress is proportional to strain but varies due to each...
Cable Subjected to a Distributed Load
Stress Concentrations in Circular Shafts
Method of Joints: Problem Solving I
Mechanical Characteristics of Steel
The tension test is fundamental for determining tensile strength. In this test, a steel specimen is stretched using a gripping device until it breaks. The data collected during this test are used...