Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Experimental small fields output factors determination for an MR-linac according to the measuring position and
José Alejandro Rojas-López1,2, Alexis Cabrera-Santiago3,4, Albin Ariel García-Andino5
1Hospital Angeles Puebla, Av. Kepler 2143, Reserva Territorial Atlixcáyotl, 72190, Puebla, Mexico.
Abstract:
Purpose. To investigate the effect of the position and orientation of the detector and its influence on the determination of output factors (OF) for small fields for a linear accelerator (MR-linac) integrated with 1.5 T magnetic resonance following the TRS-483 formalism.Methods. OF were measured for small fields in the central axis following the recommendations of the manufacturer and at the dose maximum following the TRS-483 formalism. OF were determined using a microDiamond (MD), a Semiflex (SF) 31021 ionization chamber, Gafchromic EBT3 film and were calculated in Monaco treatment planning system (TPS). Additionally, the orientation response of SF was evaluated, placing it in parallel and perpendicular direction to the radiation beam. The values were compared taking film measurements as reference. The corrected factors,ΩQclinical,msrfclinical,msr, required the use of output correction factorkQclinical,msrfclinical,msrtaken from previous reports. Finally, there are proposed experimentalkQclinical,msrfclinical,msrfor SF and MD, following the measured values in this work.Results. In fields smaller than 4 cm, the positioning of the SF and MD in the central axis or at the point of dose maximum affects the reading significantly with differences of up to 6% and 4%, respectively. For the data calculated in the TPS, the maximum difference of the OF between MD and TPS for fields greater than 2 cm was 0.6% and below this field size the TPS underestimates the OF up to 10.6%. The orientation (parallel or perpendicular) of the SF regarding the radiation beam has a considerable impact on the OF for fields smaller than 3 cm, showing a variation up to 10% for the field of 0.5 cm.Conclusion. This study provides valuable information on the challenges and limitations of measuring output factors in small fields. The outcomes have important implications for the practice of radiosurgery, underscoring the need for accuracy in detector placement and orientation, as well as the importance of using more advanced technologies and more robust measurement methods.
More Related Videos
10:22MRM Microcoil Performance Calibration and Usage Demonstrated on Medicago truncatula Roots at 22 T
Published on: January 16, 2021
14:11Quantification of Hydrogen Concentrations in Surface and Interface Layers and Bulk Materials through Depth Profiling with Nuclear Reaction Analysis
Published on: March 29, 2016
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
Mass Analyzers: Common Types