Related Experiment Video
Updated: Jun 4, 2025

Flexible Colonoscopy in Mice to Evaluate the Severity of Colitis and Colorectal Tumors Using a Validated Endoscopic Scoring System
Published on: October 16, 2013
Validation of Artificial Intelligence Computer-Aided Detection of Colonic Neoplasm in Colonoscopy
Hannah Lee1, Jun-Won Chung1, Kyoung Oh Kim1
1Division of Gastroenterology, Department of Internal Medicine, Gachon University Gil Medical Center, Incheon 21565, Republic of Korea.
Background/Objectives:
Controlling colonoscopic quality is important in the detection of colon polyps during colonoscopy as it reduces the overall long-term colorectal cancer risk. Artificial intelligence has recently been introduced in various medical fields. In this study, we aimed to validate a previously developed artificial intelligence (AI) computer-aided detection (CADe) algorithm called ALPHAON® and compare outcomes with previous studies that showed that AI outperformed and assisted endoscopists of diverse levels of expertise in detecting colon polyps.
Methods:
We used the retrospective data of 500 still images, including 100 polyp images and 400 healthy colon images. In addition, we validated the CADe algorithm and compared its diagnostic performance with that of two expert endoscopists and six trainees from Gachon University Gil Medical Center. After a washing-out period of over 2 weeks, endoscopists performed polyp detection on the same dataset with the assistance of ALPHAON®.
Results:
The CADe algorithm presented a high capability in detecting colon polyps, with an accuracy of 0.97 (95% CI: 0.96 to 0.99), sensitivity of 0.91 (95% CI: 0.85 to 0.97), specificity of 0.99 (95% CI: 0.97 to 0.99), and AUC of 0.967. When evaluating and comparing the polyp detection ability of ALPHAON® with that of endoscopists with different levels of expertise (regarding years of endoscopic experience), it was found that ALPHAON® outperformed the experts in accuracy (0.97, 95% CI: 0.96 to 0.99), sensitivity (0.91, 95% CI: 0.85 to 0.97), and specificity (0.99, 95% CI: 0.97 to 0.99). After a washing-out period of over 2 weeks, the overall capability significantly improved for both experts and trainees with the assistance of ALPHAON®.
Conclusions:
The high performance of the CADe algorithm system in colon polyp detection during colonoscopy was verified. The sensitivity of ALPHAON® led to it outperforming the experts, and it demonstrated the ability to enhance the polyp detection ability of both experts and trainees, which suggests a significant possibility of ALPHAON® being able to provide endoscopic assistance.
Related Concept Videos
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Endoscopic Procedures II: Colonoscopy