Related Experiment Video
Updated: May 6, 2026

Detection of Exosomal Biomarker by Electric Field-induced Release and Measurement EFIRM
Published on: January 23, 2015
Development of a validated novel bead extraction method for the detection of anti-PEG antibodies in human serum
William T Williams1, Kathryn Lindley1, Hong Liao1
1BioAgilytix Labs, Durham, NC, USA.
Aims:
Polyethylene glycol (PEG) is used in many applications including drug development. Due to exposure to environmental products, there is a high prevalence of preexisting anti-PEG antibodies in the global human population. The presence of anti-PEG antibodies is a concern for potentially reducing the efficacy of therapeutics after administration and represents a risk of safety events after exposure to PEGylated drug products. We developed and validated a creative and sensitive method for the detection of anti-PEG antibodies in human serum to support clinical programs for PEGylated drugs.
Methods:
In this method, biotin-PEG streptavidin beads were used to extract anti-PEG antibodies from human serum for analysis in an anti-PEG ELISA assay. The same serum sample was analyzed in an anti-drug antibody assay.
Results:
The anti-PEG antibody assay was validated with a screening cut point of 1.41 normalized signal, confirmatory cut point of 32.2% inhibition, sensitivity of 7.81 ng/mL and sufficient reproducibility, selectivity, and drug tolerance in accordance with the FDA 2019 Immunogenicity guidance.
Conclusion:
This method of removal of anti-PEG antibodies enables the use of a single sample to detect anti-drug and anti-PEG antibodies to support drug development programs.
More Related Videos
08:26Development and Validation of an Ultrasensitive Single Molecule Array Digital Enzyme-linked Immunosorbent Assay for Human Interferon-α
Published on: June 14, 2018
05:25Author Spotlight: Expanding the Scope of Multiplex Immunoassays for Lyme Borreliosis Diagnostics and Pathogen Research
Published on: July 14, 2023