Photon counting CT versus energy-integrating CT: A comparative evaluation of advances in image resolution, noise, and
Björn Heismann1, Björn Kreisler2, Robert Fasbender2
1Friedrich-Alexander-University of Erlangen-Nuremberg, Siemens Healthineers AG, Forchheim, Germany.
Background:
Photon counting computed tomography (PCCT) employs direct and spectrally resolved counting of individual x-ray quanta, enhancing image quality compared to the standard energy-integrating CT (EICT).
Purpose:
To evaluate the quantitative improvements in CT image quality metrics by comparing the first medical PCCT with a state-of-the-art EICT.
Methods:
The PCCT versus EICT noise improvement ratio R was derived from the quantum statistics of the measurement process and measured across the clinical x-ray flux range for both systems. Detector and system modulation transfer functions (MTFs) were obtained using tilted-slit and wire phantom measurements. Image root mean square (RMS) noise, noise power spectrum (NPS), and x-ray patient dose were compared using a CatPhan phantom at two identical clinical target resolutions.
Results:
The measurement of the PCCT noise improvement ratio R showed an elimination of electronic noise and a 10% noise transfer advantage. The PCCT detector MTF exhibited 3x higher angular resolution limits in comparison to EICT and close to ideal sinc behavior due to the electromagnetic formation of pixels in the PCCT semiconductor detector. This translated to 3.5x enhancements in CT system MTF ratios at 10 LP/cm, reflecting a significant improvement in millimeter range CT imaging. Both the improved quantum detection and the system MTF ratio improvement contribute to the measured 3x enhancements in image NPS at 10 LP/cm for identical image target resolution. An improvement of up to 1.7x in RMS image noise was observed accordingly. For low and ultra-low dose imaging with image filtering, dose efficiency increased between 2x and 10x, demonstrating the PCCT's capability to advance CT ultra-low dose imaging.
Conclusion:
The direct counting detection in PCCT has been shown to significantly improve sinogram noise and detector MTF ratios compared to energy integrating EICT. The observed translations into CT system MTF, image NPS, image noise, and dose ratios reflect a paradigm shift for CT image quality and dose efficiency.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System V: CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...