Broken semiflexible intraocular lens implant
American Journal of Ophthalmology
|February 15, 1985
Summary
A Kelman Quadraflex intraocular lens haptic fractured within a patient's eye, causing corneal edema and endothelial cell loss. Surgical intervention was necessary to restore vision.
Area of Science:
- Ophthalmology
- Biomedical Engineering
Background:
- Intraocular lenses (IOLs) are commonly implanted during cataract surgery.
- Semiflexible anterior chamber IOLs, such as the Kelman Quadraflex, have been used in specific clinical situations.
Observation:
- A 70-year-old male patient experienced spontaneous haptic fracture of an implanted Kelman Quadraflex IOL within the anterior chamber.
- Clinical signs included progressive inferior corneal edema and a significant, rapid decline in corneal endothelial cell density.
- The fractured haptic segment was observed to be free-floating within the anterior chamber.
Findings:
- Haptic fracture of the Kelman Quadraflex IOL occurred without apparent history of trauma.
- The free-floating haptic segment induced significant corneal edema and endothelial cell loss.
- Surgical management involved haptic and IOL removal, penetrating keratoplasty, and vitrectomy.
Implications:
- This case highlights a rare complication of IOL material failure, specifically haptic fracture.
- Corneal edema and endothelial decompensation are potential consequences of free-floating intraocular fragments.
- Complex surgical management may be required to address such complications and preserve visual function.