On the origin of MTF reduction in grating-based x-ray differential phase contrast CT imaging
Yuhang Tan1, Jiecheng Yang1, Hairong Zheng2,3,4
1Research Center for Advanced Detection Materials and Medical Imaging Devices, Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences, Shenzhen, Guangdong, China.
Background:
The complementary absorption contrast CT (ACT) and differential phase contrast CT (DPCT) can be generated simultaneously from an x-ray computed tomography (CT) imaging system incorporated with grating interferometer. However, it has been reported that ACT images exhibit better spatial resolution than DPCT images. By far, the primary cause of such discrepancy remains unclear.
Purpose:
The purpose of this study is to investigate the underlying cause of the resolution discrepancy between ACT and DPCT in a grating interferometer CT imaging system.
Methods:
In this study, theoretical derivations were performed with a -phase Talbot-Lau grating interferometer system to model the signal formation mechanism of absorption imaging and phase imaging, respectively. In addition, physical, and numerical experiments were conducted to verify the theoretical findings and assess the resolution discrepancy between ACT and DPCT under various conditions. Herein, the ACT and DPCT images were reconstructed from the filtered-back-projection algorithm using a standard Ramp filter and a standard Hilbert filter, respectively.
Results:
Experiments demonstrated that the spatial resolution of ACT and DPCT images are primarily impacted by the beam diffraction induced signal splitting. In particular, lower modulation transfer function (MTF) was observed for DPCT than ACT due to the opposite-superposition of phase signals. In addition, factors such as focal spot size, beam spectra, object composition, sample size, and detector pixel size were found to have minor impacts on the MTFs of both ACT and DPCT.
Conclusions:
In conclusion, this study reveals that the opposite-superposition of split phase signals causes the spatial resolution reduction in DPCT imaging.
More Related Videos
09:00Visualization of Failure and the Associated Grain-Scale Mechanical Behavior of Granular Soils under Shear using Synchrotron X-Ray Micro-Tomography
Published on: September 29, 2019
08:09Microfocus X-ray CT microCT Imaging of Actinia equina Cnidaria, Harmothoe sp. Annelida, and Xenoturbella japonica Xenacoelomorpha
Published on: August 6, 2019
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
X-ray Imaging