Accuracy of the Imageless Mode of the ROSA Robotic System for Targeted Resection Thickness in Total Knee
Zakareya Gamie1,2, Eustathios Kenanidis1,2,3, Georgios Douvlis1,2
1Tsiridis Orthopaedic Institute, ICAROS Clinic, Thessaloniki, Greece.
Background:
We investigated the accuracy of targeted resection thickness in patients undergoing primary Total Knee Arthroplasty (TKA) using the ROSA robotic system.
Methods:
Calliper measurements of the distal femur (DF), proximal tibia (PT), and posterior condyles (PC) were taken in 44 patients from June 2023 to January 2024.
Results:
Planned and actual resection depth difference was 0.67 mm ± 0.6 mm (mean ± SD) (p = 0.217) and 0.94 mm ± 1.15 mm (p = 0.4) for medial and lateral DF, 0.93 mm ± 0.81 mm (p = 0.001) and 0.89 mm ± 0.8 mm (p = 0.008) for medial and lateral PT, and 1.1 mm ± 0.97 mm (p = 0.001) and 1.04 mm ± 0.79 mm (p = 0.001) for medial and lateral PC, respectively.
Conclusion:
The ROSA robotic system can achieve a high degree of accuracy for planned resection thickness. Results are valid only for the imageless ROSA TKA in patients with primary knee osteoarthritis.