Background And Purpose:
Conventionally, the quality of radiotherapy treatment plans is assessed through visual inspection of dose distributions and dose-volume histograms. This study developed a framework to evaluate plan quality using dose, complexity, and robustness metrics. Additionally, a method for predicting plan robustness metrics using dose and complexity metrics was introduced for cases where plan robustness evaluation is unavailable or impractical.
Materials And Methods:
The framework and prediction models were developed and validated using 103-bronchial Volumetric Modulated Arc Therapy (VMAT)-plans. The application of the framework was demonstrated using 25-VMAT-plans. To identify significant metrics for plan evaluation, 122-metrics were analysed and narrowed down using multivariate Spearman correlation. Metric limits were set with Statistical process control (SPC). Robustness metrics were predicted using multivariable or single linear regression models based on dose-and complexity-metrics.
Results:
Twenty-five-metrics were selected based on the amount and strength of correlations. R95(dose coverage) and HI95/5(homogeneity index) stood out among the dose-metrics, while the complexity-metrics showed similar correlations. Average scenarios dose at 95 % Clinical Target Volume D95mean(CTV) and Errorbar-based Volume-Histograms (EVH) were notable for robustness metrics. Approximately 99 % of evaluated metrics fell within established SPC limits. The prediction model for D95mean(CTV) showed good performance (adjusted R2 = 0.88, mean squared error (MSE) = 3.84 × 10-6), while the model for EVH demonstrated moderate reliability (adjusted R2 = 0.52, MSE = 0.2). No statistically significant differences were found between the predicted (using dose-and complexity-metrics) and calculated robustness metrics (EVH (p-value = 0.9) and D95mean(CTV) (p-value = 1)).
Conclusions:
The developed framework enables early detection of sub-optimal, complex and non-robust treatment plans. The predictive model can be used when robustness evaluations are impractical.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...