Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
Planning transcatheter pulmonary valve implantation in the dysfunctional native RVOT: A semi-automated pipeline for
Francesco Sturla1, Alessandro Caimi2, Luca Giugno3
13D and Computer Simulation Laboratory, IRCCS Policlinico San Donato, San Donato Milanese, Italy; Department of Electronics, Information and Bioengineering, Politecnico di Milano, Milano, Italy.
Background And Objective:
Dysfunction of the right ventricular outflow tract (RVOT) is a common long-term complication following surgical repair in patients with congenital heart disease. Transcatheter pulmonary valve implantation (TPVI) offers a viable alternative to surgical pulmonary valve replacement (SPVR) for treating pulmonary regurgitation but not all RVOT anatomies are suitable for TPVI. To identify a suitable landing zone (LZ) for TPVI, three-dimensional multiphase (4D) computed tomography (CT) is used to evaluate the size, shape, and dynamic behavior of the RVOT throughout the cardiac cycle. However, manually extracting measurements from multiplanar CT reformats is operator-dependent and time-consuming. Leveraging an optical-flow (OF) algorithm, we proposed a novel semi-automated pipeline for dynamic and comprehensive geometrical analysis of the RVOT anatomy.
Methods:
Upon 4D-CT availability, at a pre-defined reference time-point, the patient-specific anatomy is semi-automatically segmented to generate the corresponding three-dimensional surface, which is navigated through a graphical user interface to define the mid-section of the potential LZ. Based on the axial length of the intended device, the proximal and distal LZ cross-sections are automatically identified. An OF-based algorithm is used to track the three LZ cross-sections frame by frame throughout the cardiac cycle, taking RVOT out-of-plane motion into account to update RVOT contours on each cross-section and to elaborate LZ geometrical changes. Finally, LZ time-dependent geometrical features are quantified and extracted.
Results:
The pipeline was successfully applied to a retrospective cohort of patients, with OF-based tracking reporting excellent agreement (r2 = 0.99) compared to manual processing, with a bias < 1% for both LZ area and perimeter, while also significantly improving time efficiency. CT-derived measurements extracted from LZ mid-section were the most influential covariates affecting the likelihood of TPVI feasibility. Among these, the minimum perimeter outperformed all other geometric LZ parameters in classifying patients as suitable for either TPVI or SPVR and achieved the highest area under the curve of 0.99, with accuracy and precision of 0.93 and 0.92, respectively.
Conclusions:
Dynamic OF-based quantification of key RVOT geometric parameters can enhance and expedite the selection process for TPVI candidates and guide optimal valve sizing during TPVI planning.