High-Sensitivity Flow Cytometry for the Reliable Detection of Measurable Residual Disease in Hematological
María Beatriz Álvarez Flores1, María Sopeña Corvinos2, Raquel Guillén Santos2
1Centro Nacional de Investigaciones Cardiovasculares Carlos III, 28029 Madrid, Spain.
Background:
Monitoring of measurable residual disease (MRD) requires highly sensitive flow cytometry protocols to provide an accurate prediction of shorter progression-free survival. High assay sensitivity generally requires rapid processing to avoid cell loss from small bone marrow sample volumes, but this requirement conflicts with the need in most clinical cytometry laboratories for long processing and acquisition times, especially when multiple MRD studies coincide on the same day.
Methods:
The proposed protocol was applied to 226 human bone marrow and 45 peripheral blood samples submitted for the study of MRD or the detection of rare cells. Samples were processed within 24 h of extraction and acquired with an eight-color flow cytometer.
Results:
The FACSLyse-Bulk protocol allows for the labelling of millions of cells in under 90 min in small sample volumes without affecting the FSC/SSC pattern or antigen expression, and it also allows antigens to be fixed to the membrane, thus avoiding the capping phenomenon.
Conclusions:
The proposed protocol would allow clinical flow cytometry laboratories to perform MRD studies in house and easily achieve a limit of detection and limit of quantification <0.001%, thus avoiding the need to outsource analysis to specialized cytometry laboratories.
More Related Videos
09:57Comprehensive Protocol to Sample and Process Bone Marrow for Measuring Measurable Residual Disease and Leukemic Stem Cells in Acute Myeloid Leukemia
Published on: March 5, 2018
11:59Detection of Residual Donor Erythroid Progenitor Cells after Hematopoietic Stem Cell Transplantation for Patients with Hemoglobinopathies
Published on: September 6, 2017