Addressing Medical Device Extractables and Leachables via Non-Target Analysis (NTA); The Analytical Evaluation
Dennis Jenke1, Piet Christiaens2, Ted Heise3
1Triad Scientific Solutions, LLC, Hawthorn Woods, IL; dennisjenke@triadscientificsolutions.com.
Medical devices release substances that can affect patient health. This study addresses challenges in analyzing these leachables using non-target analysis (NTA) and the analytical evaluation threshold (AET).
Area of Science:
- Analytical Chemistry
- Medical Device Safety
- Toxicology
Background:
- Leachables from medical devices pose potential patient health risks.
- Non-target analysis (NTA) using chromatography-mass spectrometry is crucial for profiling these substances.
- Variability in response factors complicates risk assessment strategies.
Purpose of the Study:
- To discuss challenges in applying the analytical evaluation threshold (AET) for leachables.
- To explore methods for accurate quantification of leachables and extractables.
- To address the balance between accuracy and protectiveness in risk assessment.
Main Methods:
- Screening of medical device extracts/leachates for organic substances.
- Utilizing non-target analysis (NTA) with chromatographic and mass spectrometric detection.
- Investigating practices for AET calculation, application, and quantification.
Main Results:
- Significant variation in chromatographic mass spectral response factors for leachables and extractables.
- Difficulties in interpreting the AET due to analytical uncertainty.
- Potential for false positive and false negative outcomes in risk assessment.
Conclusions:
- Response factor variability is a major challenge in medical device leachable analysis.
- Improved practices for AET application and quantification are needed.
- Balancing accuracy and protectiveness is key for reliable toxicological risk assessment.
More Related Videos
09:04Identifying Per- and Polyfluorinated Chemical Species with a Combined Targeted and Non-Targeted-Screening High-Resolution Mass Spectrometry Workflow
Published on: April 18, 2019
06:15Detection of Endotoxin in Nano-formulations Using Limulus Amoebocyte Lysate LAL Assays
Published on: January 30, 2019